The HV-3000-P, equipped with powerful HDT/VICAT testing software and high-precision LVDT sensors, offers manufacturers a reliable and efficient solution for evaluating the deflection performance of materials under specific loads and temperatures. With its multi-station design, it allows up to 6 sample to be tested simultaneously. Moreover, this instrument supports a diverse range of HDT and VICAT test options, providing flexible solutions to meet various testing needs. Its testing software also features high-precision LVDT linear calibration that ensures consistent and accurate test results.
Well-suited for diverse industries such as aerospace, material science, construction, automotive, energy, medical devices, and industrial manufacturing.
The position-adjustable LVDT probes and motorized stations enable operators to install samples and conduct tests quickly and effortlessly.
Automatically conducts correction for thermal expansion during tests to ensure the accuracy and reliability of the test results.
Employs linear calibration methods to maintain accuracy in measurements across different temperature conditions.
Utilizes a high-efficiency vortex stirrer system to achieve uniform temperature distribution in the oil bath, enhancing test repeatability and reliability.
Constructed with an enclosed extraction structure to remove fumes generated during tests, ensuring safety during operation.
The HV-3000-P is equipped with an efficient water-cooling heat exchanger and water solenoid valve to remove heat in the oil bath. Additionally, an optional and specialized water chiller is also available to cool down the oil bath more effectively.
The compact design makes the HV-3000-P an ideal instrument for laboratories and limited workspaces.
Equipped with high-precision LVDT sensor and control systems to ensure test results accuracy and reliability.
Featuring automatic operation, the HV-3000-P significantly reduces human intervention, thereby improving efficiency in the testing process.
Utilizing a reduced-sized oil bath to decrease the consumption of heat-conducting fluid, thereby improving temperature control efficiency.
With its flexible parameters setup, operators are allowed to set and modify test conditions to meet various testing requirements.
Allows operators to record and search test results to assist them in materials performance evaluation.
| Temp. range | Ambient temp. to 300℃ |
| Test station | 3 or 6 (please specify) |
| Operating system | PC based HDT software |
| Temp. control accuracy | ±0.5℃ |
| Heating rate | 2±0.2℃/min (HDT) or 50±5℃/h (VICAT) |
| Cooling method | Built in water piping (standard), water chiller (optional) |
| Cool down time (water piping) | 300 to 35℃ within 40min (when water temp. is below 25℃) |
| Test method | HDT and VICAT |
| Deflection / penetration measuring range | 0.001 to 2.500mm (LVDT sensor) |
| Span support | 64mm, 100mm, or 101.6mm (optional) |
| Stirring method | Vortex stirrer |
| Power | 1∮ AC 220V 50Hz (or specify) |
Villa Payung Indah B 27, Jawa Tengah, Pudakpayung, Kec. Banyumanik, Kota Semarang, Jawa Tengah 50265